TA0002.02 - Laboratory_Design.
4/8/33/457
Previous Help Item
Next HelpItem
Description
Comment
The Andrology Laboratory should contain adequate work space for the receipt of specimens, the holding of specimens (shaker or mixer, preferably at 37C), a microscope and hand counter, a separate bench for at least on calibrated centrifuge (so as not to cause vibrations when using the microscope), a clean writing area or computer space and an area for disposal of waste and sharp material. If samples are prepared for clinical use such as IUI or IVF, a laminar flow work space should be considered preferably in a isolated or separate room. There must be sink within easy reach with safety eye wash, suitable antiseptic hand wash cleaner and emergency bleach for spills. There needs to be a suitable work bench space of at least 2 meters for antisperm antibody assessment and for staining for sperm morphology. Given the volatile nature of some staining solutions, this should be either in a fume hood or a well ventilated room with suitable exhaust.